Title: 雙功能膝關節鬆弛度量測儀之研發
The Development of a Dual-Function Knee Ligament Arthrometer
Authors: Chun-Yen ko
柯雋彥
Contributors: NTOU:Department of Mechanical and Mechatronic Engineering
國立臺灣海洋大學:機械與機電工程學系
Keywords: 膝關節韌帶;韌帶鬆弛度測試;前/後十字韌帶;KT-1000;內/外側副韌帶;資料重現性
Knee ligament;Arthrometer;Anterior cruciate ligament;Posterior cruciate ligament;data reproducibility;KT-1000
Date: 2011
Issue Date: 2011-11-23T03:16:02Z
Abstract: 本論文研發與設計出一套新型膝關節韌帶鬆弛度量測儀,此儀器不僅可提供小腿相對大腿之前後向鬆弛度測試,亦能量測其內外偏轉測試。換而言之,雙功能膝關節韌帶鬆弛度量測儀可量測前/後十字韌帶及內/外側副韌帶鬆弛度測試。針對上述問題,我們以2-UPU並聯式機構做為膝關節鬆弛度量測儀的主體構造,2-UPU並聯式機構之自由度為4,符合機構設計之運動需求。 本論文中分別進行新型膝關節韌帶鬆弛度量測儀之重現性評估實驗,並與傳統膝關節韌帶量測儀器KT-1000進行量測數據之比較。由實驗結果可發現,新型膝關節韌帶鬆弛度量測儀擁有較佳之數據重現性;在前十字韌帶鬆弛度量測中,量測數據之標準差較KT-1000小,較不易受小腿肌肉鬆弛變量影響,可提供較為準確之量測數據。新型膝關節韌帶鬆弛度量測儀器較不易受操作人員經驗影響,將能提供骨科醫師更可靠的臨床診斷資訊。
The objective of this thesis is to develop and design a new knee ligament arthrometer; The arthrometer can measure not only the anterior/posterior (A/P) translation but also the valgus/varus angulation of the tibia. That is, the dual-function of the knee laxity test device includes the ACL/PCL test and the MCL/LCL test. Considering the above requirements, we have selected the 2-UPU parallel manipulator as the topological structure of the new knee ligament arthrometer; The 2-UPU spatial parallel manipulator has four degree-of-freedom, which satisfies our requirements. In the experiment, we assessed and verified arthrometer’s data reproducibility, and compared the measuring data of the dual function arthrometer with that of the traditional arthrometer KT-1000. From the experiment, it is confirmed that the new arthometer has a better data reproducibility. For the anterior cruciate ligament laxity test, the standard deviation of the new arthrometer’s data is smaller than that of the KT-1000,indicating that the new arthrometer will not be influenced by the muscle relaxtion of the leg and can provide a better data reliability. The new arthrometer is independent of the operator’s experience . It can provide the more reliable information for the orthopedists in the clinical diagnosis.
URI: http://ethesys.lib.ntou.edu.tw/cdrfb3/record/#G0M98720063
http://ntour.ntou.edu.tw/handle/987654321/29772
Appears in Collections:[機械與機電工程學系] 博碩士論文

Files in This Item:

File Description SizeFormat
index.html0KbHTML227View/Open


All items in NTOUR are protected by copyright, with all rights reserved.

 


著作權政策宣告: 本網站之內容為國立臺灣海洋大學所收錄之機構典藏,無償提供學術研究與公眾教育等公益性使用,請合理使用本網站之內容,以尊重著作權人之權益。
網站維護: 海大圖資處 圖書系統組
DSpace Software Copyright © 2002-2004  MIT &  Hewlett-Packard  /   Enhanced by   NTU Library IR team Copyright ©   - Feedback